Case Study

Chlorine Determination in the Analysis of Coal and Coke

Sample Preparation
Samples may be analyzed as air dried
5E-FL2350 Fluorine/Chlorine Analyzer

Instrument
5E-FL2350 Fluorine/Chlorine Analyzer

Instrument Settings
Oxygen : Flow 500ml/min; purity 99.5%            Vapour :2ml/min
Main Furnace Temperature:1100 ℃                  Condensator  temp. :20℃
Calibration Standard: GBT3558-2014               Nominal Sample Mass: 100 mg
Analysis Time: 45 minutes/per test

Special Modifications to Unit

None

Method Approval
GBT3558-2014 and ISO 587 and ASTM D6721

Procedure

1. Prepare instrument as outlined in the operator’s instruction manual (check the solutions , pure water content and electrodes ,oxygen gas flow and etc.).
2. Do three blanks sample “Clt” titration (initialize the titration system)
3. Combustion pretreatment with two empty crucibles (this process only for first time of using of each day)
4. weigh a portion(0.5g) of solid waste sample in crucible, then cover the sample with a flat layer of silica sand;
5. Load two crucibles with sample as prepared in step4 , then” start” analysis.
6. 2 samples of calibration substance to check the calibration after 10 duplicate sample analysis;
7. Repetition of point 4 to point 6 until all samples are processed;

Note
Do Calibrating refer to operator’s instruction manual for details.

Typical Results

CodeNameMass(g)Clad(%)RangeAverageClt, ad(%)
1Coal0.50450.210.040.210.21
2Coal0.50370.23
3Coal0.50310.22
4Coal0.50900.19